Thin layer chromatography analysis of PAHs from water
Thin layer chromatography analysis of PAHs from water
Thin layer chromatography analysis of PAHs from water
- Application Number
- 401070
- Method
- TLC
- Results
- Substances
- naphtalene; acenaphthylene; acenaphthene; fluorene; phenanthrene; anthracene; pyrene; benzanthracene; chrysene; benzofluoranthene; fluoranthene; benzopyrene; indenopyrene; dibenzanthracene; benzoperylene
- Product(s)
Phase REF Webshop Nano-SIL-PAH
811051
- Matrix
- water
- Sample(s)
- Conditions
Eluent: dichloromethane, chamber cooled to - 20 °C for 30 min prior to development Migration distance: 9 cm in 20 min at - 20 °C
- Detection
- see Note
- Note
- Detection: dip dried plate for 2 seconds in a solution of paraffin / n-hexane (1:2, v/v), dry again; native fluorescence (dipping in paraffin stabilizes and enhances the fluorecence intensity) Calibration function of individual PAHs HPTLC separation of 6 PAHs acc. to German drinking water specifications
- Author
- Möller K.
- Source
- GIT Spezial Chromatographie 2/94, p. 81 - 84
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.
- Dateien
- Lädt...Lädt...