Thin layer chromatography analysis of PAHs from water

Thin layer chromatography analysis of PAHs from water

Thin layer chromatography analysis of PAHs from water

Application Number
401070
Method
TLC
Results


Substances
naphtalene; acenaphthylene; acenaphthene; fluorene; phenanthrene; anthracene; pyrene; benzanthracene; chrysene; benzofluoranthene; fluoranthene; benzopyrene; indenopyrene; dibenzanthracene; benzoperylene
Product(s)
PhaseREFWebshop

Nano-SIL-PAH

811051

Matrix
water
Sample(s)
Conditions

Eluent: dichloromethane, chamber cooled to - 20 °C for 30 min prior to development Migration distance: 9 cm in 20 min at - 20 °C

Detection
see Note
Note
Detection: dip dried plate for 2 seconds in a solution of paraffin / n-hexane (1:2, v/v), dry again; native fluorescence (dipping in paraffin stabilizes and enhances the fluorecence intensity) Calibration function of individual PAHs HPTLC separation of 6 PAHs acc. to German drinking water specifications
Author
Möller K.
Source
GIT Spezial Chromatographie 2/94, p. 81 - 84
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.