Standard SPE procedure for CHROMABOND Carbon A

Standard SPE procedure for CHROMABOND Carbon A

Standard SPE procedure for CHROMABOND Carbon A

Application Number
306130
Method
SPE
Results
Substances
Product(s)
PhaseREFWebshop

CHROMABOND Carbon A

730167

Matrix
-
Sample(s)
Conditions

Column conditioning: 3 x 6 mL methanol, then 3 x 6 mL water

Sample aspiration: force or aspirate sample at a flow from 5-20 mL/min

Washing: 1-10 mL water

Drying: 5-10 min vacuum oder nitrogen flow

Elution: 3 x 2 mL methanol

Solvent change: depending on further analysis

Detection
-
Subsequent analysis
Note
Author
Application department
Source
MACHEREY-NAGEL, Germany 2016
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.