Standard SPE procedure for CHROMABOND HR-XC
Standard SPE procedure for CHROMABOND HR-XC
Standard SPE procedure for CHROMABOND HR-XC
- Application Number
- 304790
- Method
- SPE
- Results
- Substances
- Product(s)
Phase REF Webshop CHROMABOND HR-XC
730952
- Matrix
- -
- Sample(s)
Individual sample preparation in reference to the compounds and matrix (Adjust pH value if necessary)
- Conditions
Column conditioning: 5 mL methanol, then 5 mL water (Do not let run the column dry!)
Sample aspiration: The prepared sample is passed through the column by vacuum or pressure
Washing 1: 2 mL 0.1 M HCl in water
Washing 2 / Elution 1: 2 mL methanol (elution of neutral and acidic compounds)
Drying: With nitrogen or air
Elution 2: 5 mL methanol / 5 % NH3 (elution of basic compounds)
Evaporation and reconstitution (if necessary)
- Detection
- -
- Subsequent analysis
- Note
- The volumes are typical for columns 3 mL/200 mg. The method can be easily adjusted to different column sizes by changing the volumes in the single steps. Depending on the application the solvent for the elution can be changed and may give improved results.
- Author
- SPE department
- Source
- MACHEREY-NAGEL, Germany 2008
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.