Codeine from serum or urine on CHROMABOND HR-XC

Codeine from serum or urine on CHROMABOND HR-XC

Codeine from serum or urine on CHROMABOND HR-XC

Application Number
304770
Method
SPE
Results
Recovery rates of the extraction from serum in comparison with competitors

Title: Recovery rates of the extraction from serum in comparison with competitors

Recovery rate of the extraction from urine on CHROMABOND HR-XC

Title: Recovery rate of the extraction from urine on CHROMABOND HR-XC

Substances
codeine
Product(s)
PhaseREFWebshop

CHROMABOND HR-XC

730952

Matrix
serum; urine
Sample(s)

1 mL spiked serum or urine (c: 1 µg/mL) + 200 µL 2 % H3PO4

Conditions

Column conditioning: 5 mL methanol, then 5 mL water (Do not let run the column dry!)

Sample aspiration: The prepared sample is passed through the column by vacuum or pressure.

Washing: With the following washing mixtures the impurities are removed: a) 2 mL 0.1 N HCl b) 2 mL methanol

Drying: With nitrogen or air

Elution: Analytes are eluted with 3 x 2 mL methanol / 5 % NH3.

Evaporation and reconstitution with 1 mL of mobile phase from subsequent HPLC.

Detection
-
Subsequent analysis
HPLC determination of the recovery rates with EC 125/4 NUCLEODUR C18 Gravity, 5 µm (REF 760100.40) in reference to Appl. No. 122210. Please consider also Appl. No. 117950 "Separation of alkaloids with NUCLEODUR".
Note
Author
SPE department
Source
MACHEREY-NAGEL (2008)
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.