Codeine from serum or urine on CHROMABOND HR-XC
Codeine from serum or urine on CHROMABOND HR-XC
Codeine from serum or urine on CHROMABOND HR-XC
- Application Number
- 304770
- Method
- SPE
- Results
Title: Recovery rates of the extraction from serum in comparison with competitors
Title: Recovery rate of the extraction from urine on CHROMABOND HR-XC
- Substances
- codeine
- Product(s)
Phase REF Webshop CHROMABOND HR-XC
730952
- Matrix
- serum; urine
- Sample(s)
1 mL spiked serum or urine (c: 1 µg/mL) + 200 µL 2 % H3PO4
- Conditions
Column conditioning: 5 mL methanol, then 5 mL water (Do not let run the column dry!)
Sample aspiration: The prepared sample is passed through the column by vacuum or pressure.
Washing: With the following washing mixtures the impurities are removed: a) 2 mL 0.1 N HCl b) 2 mL methanol
Drying: With nitrogen or air
Elution: Analytes are eluted with 3 x 2 mL methanol / 5 % NH3.
Evaporation and reconstitution with 1 mL of mobile phase from subsequent HPLC.
- Detection
- -
- Subsequent analysis
- HPLC determination of the recovery rates with EC 125/4 NUCLEODUR C18 Gravity, 5 µm (REF 760100.40) in reference to Appl. No. 122210. Please consider also Appl. No. 117950 "Separation of alkaloids with NUCLEODUR".
- Note
- Author
- SPE department
- Source
- MACHEREY-NAGEL (2008)
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.