PCB in oil samples

PCB in oil samples

PCB in oil samples

Application Number
301380
Method
SPE
Results
Substances
PCB
Product(s)
PhaseREFWebshop

CHROMABOND SiOH-H2SO4/SA and CHROMABOND SiOH

730125

Matrix
oil
Sample(s)

Extract oil-contaminated solids with n-hexane. Homogenise other oil samples and dissolve 1.5 to 2.0 g in 50 mL n-hexane. Water which may cause turbidities can be removed with sodium sulphate.

Conditions

Determination with reference to German Standard DIN 51 527

Column conditioning: Let 1 mL n-hexane flow through the CHROMABOND SiOH-H2SO4/SA column.

Sample application: Aspirate or force 500 µL sample through the CHROMABOND SiOH-H2SO4/SA column. This phase offers better removal of interfering substances due to sulphonisation. Place CHROMABOND SiOH-H2SO4/SA column on top of CHROMABOND SiOH column with the aid of an adaptor (REF 730100) and after at least 30 s flush sample into the SiOH column with 2 x 1 mL n-hexane.

Elution: Elute SiOH column with 3 x 0.5 mL n-hexane.

Evaporation: Adjust to a suitable concentration for subsequent GC analysis by evaporation of n-hexane in a stream of nitrogen or by dilution with n-hexane.

Detection
-
Subsequent analysis
Note
Recovery rates [%]: PCB 28 99 PCB 52 95 PCB 101 99 PCB 138 94 PCB 153 99 PCB 180 96 PCB 209 101
Author
SPE department
Source
MACHEREY-NAGEL
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.