Analysis of PFAS in aqueous samples by SPE and LC-MS/MS according to EPA Method 1633
Analysis of PFAS in aqueous samples by SPE and LC-MS/MS according to EPA Method 1633
Analysis of PFAS in aqueous samples by SPE and LC-MS/MS according to EPA Method 1633
- Application Number
- 129620
- Method
- HPLC
- Results
- Substances
- PFAS, see recovery rates
- Product(s)
Phase REF Webshop NUCLEODUR® Phenyl-Hexyl
760576.20
NUCLEODUR® Phenyl-Hexyl
760573.20
- Matrix
- n aqueous, solid (soil, biosolids, sediment) and tissue samples
- Sample(s)
Sample preparation routine is found in application database entry 307040.
- Conditions
DELAY Column EC 50/2 NUCLEODUR® Phenyl-Hexyl (REF 760573.20) Column EC 100/2 NUCLEODUR® Phenyl-Hexyl, 3 µm (REF 760576.20) Eluent A 5 mM ammonium acetate in water Eluent B 5 mM ammonium acetate in methanol Gradient Hold 40 % B for 1 min, in 8 min from 40 % B to 95 % B,hold 95 % B for 3 min, in 0.1 min to 40 % B, hold 40 % B for 2.9 min Flow rate 0.3 mL/min Temperature 40 °C Injection volume 2 µL
- Detection
- LC-MS/MS
- Note
- For detailed sample preparation routine, please visit entry 307040. For further information, please download the application note „Analysis of PFAS in aqueous samples by SPE and LC-MS/MS according to EPA Method 1633“: https://www.mn-net.com/chromatography-downloads
- Author
- Application Department
- Source
- Macherey-Nagel, 2025
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.