PFAS analysis in water matrices according to EPA 8327 - incl. NUCLEODUR PFAS Delay HPLC column

PFAS analysis in water matrices according to EPA 8327 - incl. NUCLEODUR PFAS Delay HPLC column

PFAS analysis in water matrices according to EPA 8327 - incl. NUCLEODUR PFAS Delay HPLC column

Application Number
129060
Method
HPLC
Results
null

Legend: Chromatograms of PFAS spiked samples according to EPA 8327 on NUCLEODUR PFAS - A: β = 10.0 ng/L for each compound - B: β = 25.0 ng/L for each compound - C: β = 50.0 ng/L for each compound - D: β = 100.0 ng/L for each compound

null

Legend: Impeding of PFOA instrument contamination with a NUCLEODUR PFAS Delay column

null

Legend: Recovery rates of PFAS according to EPA 8327 method for drinking water

null

Substances
PFAS, see retention times
Product(s)
PhaseREFWebshop

NUCLEODUR PFAS

760663.20

NUCLEODUR PFAS Delay

760673.20

Screw closure, N 9, PP

702402

Screw neck vial, N 9, PP

702009

Matrix
water
Sample(s)

Sample pretreatment for water samples according to method 3512.

Sample dilution

  • Dilute sample 50:50 with MeOH and 0.1 % acetic acid, spiked with isotopically labeled surrogates
  • Shake for 2 min
  • Fill up sample solution in polypropylene vials with fluorine-free polyimide septum.
Conditions
  • Eluent A: 5 mM ammonium acetate in water
  • Eluent B: 5 mM ammonium acetate in methanol
  • Gradient: in 4.5 min from 5.0 % B to 95 % B, hold 95 % B for 0.5 min, in 0.05 min to 5.0 % B, hold 5.0 % B for 1.45 min
  • Flow rate: 0.33 mL/min
  • Temperature: 40 °C
  • Injection volume: 20 μL
Detection
MS/MS
Note
This application is based on EPA 8327. A direct injection method for PFAS in water. MS conditions: AB Sciex QTRAP 5500 - Acquisition mode: SRM - Interface: ESI - Polarity: negative - Curtain gas: 30 psig - Collision gas: medium - Ion spray voltage: – 4500 V - Temperature: 400 °C - Ion source gas 1: 50 psig - Ion source gas 2: 60 psig - Detection window: 60 s For further information, check out: www.mn-net.com/pfas
Author
Application department
Source
MACHEREY-NAGEL, Germany 2021
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.