Analysis of pesticides on core-shell phase with LC-MS/MS detection

Analysis of pesticides on core-shell phase with LC-MS/MS detection

Analysis of pesticides on core-shell phase with LC-MS/MS detection

Application Number
128150
Method
HPLC
Results


Substances
acetamiprid; atrazine; azoxystrobin; carbaryl; chlortoluron; chlorpyrifos; clofentezine; clothianidin; coumaphos; cyanazine; desethylatrazine; desisopropylatrazine; diazinon; difenoconazole; diuron; ethoprophos; hexazinone; imazalil; imidacloprid; isoproturon; linuron; methabenzthiazuron; methomyl; metobromuron; metolachlor; monolinuron; myclobutanil; phosalone; piperonylbutoxide; propazine; propyzamide; terbuthylazine; thiacloprid
Product(s)
PhaseREFWebshop

NUCLEOSHELL PFP

763532.20

Matrix
Sample(s)

For an enrichment from serum by SPE see MN Appl. No. 306360 Sample in 1 mL dest. water - acetonitrile (95:5, v/v)

Conditions

Eluent A: dest. water + 0,1 formic acid Eluent B: acetonitrile + 0,1 formic acid Gradient: 5 - 95 % B in 15 min, 95 % B for 5 min, 95 - 5 % B in 1 min, 5 % B for 9 min Flow rate: 0.30 mL/min Temperature: 40 °C Injection volume: 5 µL

Detection
LC-MS/MS AB Sciex API 5500 Ion source: Turbo Spray (ESI) Scan type: SRM Polarity: positive Curtain gas: 20 psig Ion spray voltage: 5500 V Temperature: 450 °C Gas 1 (nebulizer): 40 psig Gas 2 (turbo gas): 40 psig Collision gas: medium
Note
Author
SPE department
Source
MACHEREY-NAGEL, Germany 2017
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.