Quantification of 25-OH vitamin D3 by online-SPE-LC-MS

Quantification of 25-OH vitamin D3 by online-SPE-LC-MS

Quantification of 25-OH vitamin D3 by online-SPE-LC-MS

Application Number
127840
Method
HPLC
Results



Substances
Product(s)
PhaseREFWebshop

NUCLEODUR C8 ec

2 x 732345.20 on request

NUCLEOSHELL PFP

763532.20

Matrix
serum
Sample(s)

Serum samples from patients are precipitated with zinc sulfate solution and an organic solvent with the addition of a deuterated internal standard. Then a centrifugation follows. Due to the high quantity of samples 96-deep-well collecting plates (e.g., REF 738650.5) are used instead of 1.5 mL standard vials. The sample pretreatment is directly carried out in the cavities of the plate.

Conditions

The enrichment of analyts as well as the removal of interfering substances are received by online solid phase extraction (online-SPE) with C8 ec extraction columns. For a faster working process two SPE extraction columns (1) + (2) are used with changing steps (s. column switching). Lipophilic substances like vitamin D3 are retained, but polar compounds like e.g., salts, proteins, amino acids, carbohydrates and water soluble vitamins can pass. After sample loading onto the SPE extraction column the analyts are eluated by changing the flow direction and an increase of the organic part of eluent. The eluent flow is diverted onto the analytical HPLC column (3) for the chromatographic separation. The analytes are ionized with the MS electrospray method and detected by tandem mass spectrography.

Eluent A: 2 mL formic acid is filled up with water (LCMS Optigrade) to 1 L volume Eluent B: 2 mL formic acid is filled up with 700 mL methanol (Optigrade) and 300 mL acetonitrle (Optigrade)

Injection volume: 5 µL / 20 µL

Gradient and flow rate:

Online-SPE (extraction) 0 to 1.1 min: A 98 % 0 to 1.1 min: flow 0.8 mL/min 1.1 to 4 min: flow 0.4 mL/min 2.7 to 3 min: A 100 % 3 to 4 min: A 2%

HPLC (analysis) 0 to 1.2 min: A 30 % 1.25 to 2.65 min: A 20 % 2.7 to 3 min: A 10 % 3.1 min: A 30 % 0 to 4 min: flow 0.4 mL/min

Oven temperature: 40 °C Rack temperature: 15 °C

Detection
see Note:
Note
Detection: LC-MS Nebulizing gas flow: 3.0 L/min Drying gas flow: 10.0 L/min Interface: ESI Interface voltage: 4.0 kV Interface current: 1.8 µA Interface temperature: 300 °C DL temperature: 250 °C Heat block temperature: 400 ° C Conversion dynode voltage: 10.0 kV Detector voltage: 1.66 kV IG vacuum: 2.1e-003 Pa PG vacuum: 9.9e+001 Pa CID gas: 270 kPa MRM (Multiple Reaction Monitoring): 25-OH vitamin D3d3 (internal standard) 386.4 → 232.3 386.4 → 257.2 25-OH vitamin D3 (analyte) 383.2 → 259.2 383.2 → 229.2 383.2 → 253.3
Author
Kröner, L.
Source
Labor Dr. Wisplinghoff, Köln, Germany
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.
Dateien