Comparing of selectivity for organic acids
Comparing of selectivity for organic acids
Comparing of selectivity for organic acids
- Application Number
- 127250
- Method
- HPLC
- Results
Legend:
1 - malic acid 2 - acetic acid 3 - citric acid- Substances
- malic acid; acetic acid; citric acid
- Product(s)
Phase REF Webshop NUCLEODUR C18 Gravity-SB
760617.40
NUCLEODUR C18 Gravity
760100.40
NUCLEODUR C18 Pyramid
760201.40
- Matrix
- -
- Sample(s)
1 mg/mL each, solved in eluent
- Conditions
- Eluent: 20 mmol/L KH2PO4, pH 2.7
- Flow rate: 0.6 mL/min
- Temperature: 20 °C
- Injection volume: 2.0 µL
- Detection
- UV, 215 nm
- Note
- Under the given conditions the highest retention time is achieved with NUCLEODUR C18 Gravity-SB. NUCLEODUR C18 Gravity was only used for illustration and not compatible with fully aqueous conditions.
- Author
- HPLC department
- Source
- MACHEREY-NAGEL, Germany 2015
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.
- Dateien