Comparison of MS-bleeding of NUCLEODUR Phenyl-Hexyl and NUCLEODUR C18 Gravity

Comparison of MS-bleeding of NUCLEODUR Phenyl-Hexyl and NUCLEODUR C18 Gravity

Comparison of MS-bleeding of NUCLEODUR Phenyl-Hexyl and NUCLEODUR C18 Gravity

Application Number
125840
Method
HPLC
Results

Substances
Product(s)
PhaseREFWebshop

NUCLEODUR Phenyl-Hexyl

760578.30

NUCLEODUR C18 Gravity

760083.30

Matrix
-
Sample(s)
Conditions
  • Eluent A: acetonitrile
  • Eluent B: water
  • Gradient: 5-95 % A in 25 min
  • Flow rate: 0.2 mL/min
  • Temperature: 25 °C
Detection
MS, ESI positive Mass range: 50 - 1000 m/z
Note
Bleeding of NUCLEODUR Phenyl-Hexyl is under given conditions comparable to NUCLEODUR C18 Gravity.
Author
HPLC department
Source
MACHEREY-NAGEL, Germany 2012
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.