Separation of Guanosine Phosphates
Separation of Guanosine Phosphates
Separation of Guanosine Phosphates
- Application Number
- 123940
- Method
- HPLC
- Results
Legend:
1 - guanosine monophosphate 2 - guanosine diphosphate 3 - guanosine triphosphate- Substances
- guanosine monophosphate; guanosine diphosphate; guanosine triphosphate
- Product(s)
Phase REF Webshop NUCLEODUR HILIC
760551.40
- Matrix
- Sample(s)
Standard samples: guanosine monophosphate: 0.5 mg/ml in eluent guanosine diphosphate: 0.5 mg/ml in eluent guanosine triphosphate: 0.5 mg/ml in eluent
Test sample: mixture from standard samples with ratio 1:2:2, v/v/v
- Conditions
Eluent: 100 mM ammonium acetate, adjusted with acetic acid to pH 5.28 / acetonitrile (30:70, v/v) Flow rate: 1.3 mL/min Temperature: 25 °C Injection volume: 4 µL
- Detection
- UV, 259 nm
- Note
- Author
- HPLC department
- Source
- MACHEREY-NAGEL, Germany 2010
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.