Separation of EFSA PAHs on NUCLEODUR C18 PAH

Separation of EFSA PAHs on NUCLEODUR C18 PAH

Separation of EFSA PAHs on NUCLEODUR C18 PAH

Application Number
123930
Method
HPLC
Results

Legend:
1 - benzo[c]fluorene 2 - cyclopenta[cd]pyrene 3 - benz[a]anthracene 4 - chrysene 5 - 5-methylchrysene 6 - benzo[j]fluoranthene 7 - benzo[b]fluoranthene 8 - benzo[k]fluoranthene 9 - benzo[a]pyrene 10 - dibenzo[a,l]pyrene 11 - dibenz[a,h]anthracene 12 - benzo[ghi]perylene 13 - indeno[1,2,3-cd]pyrene 14 - dibenzo[a,e]pyrene 15 - dibenzo[a,i]pyrene 16 - dibenzo[a,h]pyrene

Substances
benzo[c]fluorene; cyclopenta[cd]pyrene; benz[a]anthracene; chrysene; 5-methylchrysene; benzo[j]fluoranthene; benzo[b]fluoranthene; benzo[k]fluoranthene; benzo[a]pyrene; dibenzo[a,l]pyrene; dibenz[a,h]anthracene; benzo[ghi]perylene; indeno[1,2,3-cd]pyrene; dibenzo[a,e]pyrene; dibenzo[a,i]pyrene; dibenzo[a,h]pyrene
Product(s)
PhaseREFWebshop

NUCLEODUR C18 PAH

760784.40

Matrix
-
Sample(s)

EFSA PAH mixture, c=10 mg/L

Conditions
  • Eluent A: methanol / water (70:30, v/v)
  • Eluent B: acetonitrile
  • Gradient: 0-20 % B in 1.5 min, 20-50 % B in 1.5 min, 50-100 % B in 6 min, 100 % B for 13 min, 100-0 % B in 0.5 min
  • Flow rate: 1.5 mL/min
  • Temperature: 35 °C
  • Injection volume: 2 µL
Detection
UV, 251 nm
Note
Author
HPLC department
Source
MACHEREY-NAGEL, Germany 2010
Disclaimer
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.