Separation of EFSA PAHs on NUCLEODUR C18 PAH
Separation of EFSA PAHs on NUCLEODUR C18 PAH
Separation of EFSA PAHs on NUCLEODUR C18 PAH
- Application Number
- 123930
- Method
- HPLC
- Results
Legend:
1 - benzo[c]fluorene 2 - cyclopenta[cd]pyrene 3 - benz[a]anthracene 4 - chrysene 5 - 5-methylchrysene 6 - benzo[j]fluoranthene 7 - benzo[b]fluoranthene 8 - benzo[k]fluoranthene 9 - benzo[a]pyrene 10 - dibenzo[a,l]pyrene 11 - dibenz[a,h]anthracene 12 - benzo[ghi]perylene 13 - indeno[1,2,3-cd]pyrene 14 - dibenzo[a,e]pyrene 15 - dibenzo[a,i]pyrene 16 - dibenzo[a,h]pyrene- Substances
- benzo[c]fluorene; cyclopenta[cd]pyrene; benz[a]anthracene; chrysene; 5-methylchrysene; benzo[j]fluoranthene; benzo[b]fluoranthene; benzo[k]fluoranthene; benzo[a]pyrene; dibenzo[a,l]pyrene; dibenz[a,h]anthracene; benzo[ghi]perylene; indeno[1,2,3-cd]pyrene; dibenzo[a,e]pyrene; dibenzo[a,i]pyrene; dibenzo[a,h]pyrene
- Product(s)
Phase REF Webshop NUCLEODUR C18 PAH
760784.40
- Matrix
- -
- Sample(s)
EFSA PAH mixture, c=10 mg/L
- Conditions
- Eluent A: methanol / water (70:30, v/v)
- Eluent B: acetonitrile
- Gradient: 0-20 % B in 1.5 min, 20-50 % B in 1.5 min, 50-100 % B in 6 min, 100 % B for 13 min, 100-0 % B in 0.5 min
- Flow rate: 1.5 mL/min
- Temperature: 35 °C
- Injection volume: 2 µL
- Detection
- UV, 251 nm
- Note
- Author
- HPLC department
- Source
- MACHEREY-NAGEL, Germany 2010
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.
- Dateien