Determination of cannabinoids (THC) in plasma and serum samples with GC-MS
Determination of cannabinoids (THC) in plasma and serum samples with GC-MS
Determination of cannabinoids (THC) in plasma and serum samples with GC-MS
- Application Number
- 215350
- Method
- GC
- Results
- Substances
- Tetrahydrocannabinol (THC), Hydroxy Tetrahydrocannabinol (THC-OH), Carboxy Tetrahydrocannabinol (THC-COOH), D3-Tetrahydrocannabinol (THC-D₃), D3-Hydroxy Tetrahydrocannabinol (THC-OH-D₃), D3-Carboxy Tetrahydrocannabinol (THC-COOH-D₃)
- Product(s)
Phase REF Webshop OPTIMA 5 HT
726106.30
MSTFA
701270.201
- Matrix
- serum and plasma
- Sample(s)
Sample pretreatment
- Homogenize urine sample by stirring
- Fill 0.5 mL sample into a safe-lock tube
- Add standard solution, internal standard solution as described in table 1
- Add 25 µL acetic acid (25 %) and 1000 µL of a mixture of ethyl acetate – n-hexane (9+1, v+v)
- Shake vigorously for 30 sec
- Centrifuge at room temperature at 13400 rpm for 10 min
- Take up 500 µL of the organic phase for derivatization in a vial
- Concentrate organic phase at 40 °C under nitrogen stream to dryness
- Add 50 µL ethyl acetate and 50 µL MSTFA (REF 701270.201)
- Shake vigorously for 30 sec and incubate mixture at 80 °C for 30 min
- Take sample solution for GC-MS analysis
- Conditions
- Injection volume: 1 µL
- Injection mode: Splitless
- Injection temperature: 250 °C
- Carrier gas: Helium
- Flow: 1.31 mL/min
- Temperature: 70 °C [2 min] --> [20 °C/min] --> 250 °C [4 min] --> [20 °C/min] --> 300 °C [17 min]
- Detection
- MS, see Note
- Note
- MS conditions: GCMS-QP2010plus, Shimadzu - Ion source: EI - Scan type: SIM - Tune: Autotune - Ion Source temperature: 200 °C - Interface temperature: 250 °C - Solvent delay: 4 min More information about the determination of cannabinoids (THC) can be found in the brochure "Cannabis testing" at https://www.mn-net.com/highlighted-analytical-topics-in-chromatography
- Author
- Application department
- Source
- MACHEREY-NAGEL, Germany 2018
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.