Separation of xylene isomers and styrene
Separation of xylene isomers and styrene
Separation of xylene isomers and styrene
- Application Number
- 214710
- Method
- GC
- Results
Legend:
1 - p-xylene (bp 138 °C) 2 - m-xylene (bp 139 °C) 3 - o-xylene (bp 144 °C) 4 - styrene (bp 145 °C)- Substances
- p-xylene; m-xylene; o-xylene; styrene
- Product(s)
Phase REF Webshop OPTIMA FFAPplus
726241.30
- Matrix
- -
- Sample(s)
Mix of substances each 1 µL/mL in pentane
- Conditions
- Injection volume: 1.0 µL, split 1:100
- Injection temperature: 260 °C
- Carrier gas: helium, 2.0 mL/min
- Temperature program: 80 °C --> [8 °C/min] --> 150 °C [5 min]
- Detection
- FID, 260 °C
- Note
- Author
- GC department
- Source
- MACHEREY-NAGEL, Germany 2012
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.