Separation of solvents on OPTIMA FFAPplus
Separation of solvents on OPTIMA FFAPplus
Separation of solvents on OPTIMA FFAPplus
- Application Number
- 214580
- Method
- GC
- Results
Legend:
1 - n-hexane 2 - cyclohexane 3 - acetone 4 - ethyl acetate 5 - methyl ethyl ketone 6 - dichloromethane 7 - benzene 8 - methyl isobutyl ketone 9 - perchloroethylene 10 - toluene 11 - n-butyl acetate 12 - undecane 13 - ethylbenzene 14 - o-xylene 15 - styrene- Substances
- n-hexane; cyclohexane; acetone; ethyl acetate; methyl ethyl ketone; dichloromethane; benzene; methyl isobutyl ketone; perchloroethylene; toluene; n-butyl acetate; undecane; ethylbenzene; o-xylene; styrene
- Product(s)
Phase REF Webshop OPTIMA FFAPplus
726241.30
- Matrix
- -
- Sample(s)
Mix of substances
- Conditions
- Injection volume: hot needle, split 1:200
- Injection temperature: 230 °C
- Carrier gas: Helium, 1.1 bar
- Temperature: 60 °C [8min] --> [15 °C/min] --> 150 °C [10 min]
- Detection
- FID, 260 °C
- Note
- Author
- GC department
- Source
- MACHEREY-NAGEL, Germany 2012
- Disclaimer
- Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products. MACHEREY-NAGEL does not warrant the correctness of the applications.